Evaluation of the PotoClean(®) decontamination technology for reprocessing of water supply lines in dental units during routine work.

GMS Krankenhaushygiene interdisziplinar Pub Date : 2012-01-01 Epub Date: 2012-04-04 DOI:10.3205/dgkh000194
Axel Kramer, Torsten Koburger, Lisa-Dorothea Taube, Michael Menzel, Georg Meyer, Ojan Assadian
{"title":"Evaluation of the PotoClean(®) decontamination technology for reprocessing of water supply lines in dental units during routine work.","authors":"Axel Kramer,&nbsp;Torsten Koburger,&nbsp;Lisa-Dorothea Taube,&nbsp;Michael Menzel,&nbsp;Georg Meyer,&nbsp;Ojan Assadian","doi":"10.3205/dgkh000194","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>A frequent problem in dental units is the microbial contamination of water and biofilm formation in the water supply lines. After random identification of a bacterial contaminated dental unit (310 cfu/ml) in a practise with 3 dental units we implemented the present study to evaluate the efficacy of the PotoClean(®) technology, based on anodic oxidation.</p><p><strong>Method: </strong>The efficacy of a regular low concentrated permanent decontamination (1 mg Cl/L) with an additional intensive decontamination by PotoClean(®) (three times 20 mg Cl/ml for 2 h) on three dental units was tested over 7 months. Microbial contamination, total chlorine concentration and redox potential have been analyzed. Dental unit A and B was 15 years old, unit C 5 years.</p><p><strong>Results: </strong>After 3 intensive decontaminations, in dental unit A and B the number of bacteria and moulds could be reduced less than 7 d. Thereafter the bacteria counts increased again during the subsequent 7 month period and the amount of moulds was with some exceptions 300 cfu/ml, although PotoClean(®) was constantly added in the system (1 mg Cl/L). After further 7.5 month only with low concentrated permanent disinfection (1 mg Cl/L) both units were successful decontaminated. Dental unit C represented an object which was easier to decontaminate because of the advanced construction (prevention of water stagnation) and the shorter useful life. At the beginning of the decontamination it was no bacterial contamination, but moulds were contained (300 cfu/ml). Already after the first intensive decontamination, no further bacteria and moulds could be detected.</p><p><strong>Discussion: </strong>An important factor for the efficacy of PotoClean(®) was the age of the units and their construction. For a new generation of dental units PotoClean(®) was effective during the whole period of monitoring. For two old types of dental unit with massive biofilm development the successful decontamination needed more than 7 month.</p><p><strong>Conclusion: </strong>The PotoClean(®) technology has resulted in even old-type turbines with intensive biofilm formation to complete decontamination. In a recent turbine design already after the first intensive decontamination with PotoClean(®) and its continuous use (1 mg Cl/L) no more contamination by bacteria and moulds were detectable.</p>","PeriodicalId":88496,"journal":{"name":"GMS Krankenhaushygiene interdisziplinar","volume":"7 1","pages":"Doc10"},"PeriodicalIF":0.0000,"publicationDate":"2012-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3205/dgkh000194","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"GMS Krankenhaushygiene interdisziplinar","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3205/dgkh000194","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2012/4/4 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3

Abstract

Background: A frequent problem in dental units is the microbial contamination of water and biofilm formation in the water supply lines. After random identification of a bacterial contaminated dental unit (310 cfu/ml) in a practise with 3 dental units we implemented the present study to evaluate the efficacy of the PotoClean(®) technology, based on anodic oxidation.

Method: The efficacy of a regular low concentrated permanent decontamination (1 mg Cl/L) with an additional intensive decontamination by PotoClean(®) (three times 20 mg Cl/ml for 2 h) on three dental units was tested over 7 months. Microbial contamination, total chlorine concentration and redox potential have been analyzed. Dental unit A and B was 15 years old, unit C 5 years.

Results: After 3 intensive decontaminations, in dental unit A and B the number of bacteria and moulds could be reduced less than 7 d. Thereafter the bacteria counts increased again during the subsequent 7 month period and the amount of moulds was with some exceptions 300 cfu/ml, although PotoClean(®) was constantly added in the system (1 mg Cl/L). After further 7.5 month only with low concentrated permanent disinfection (1 mg Cl/L) both units were successful decontaminated. Dental unit C represented an object which was easier to decontaminate because of the advanced construction (prevention of water stagnation) and the shorter useful life. At the beginning of the decontamination it was no bacterial contamination, but moulds were contained (300 cfu/ml). Already after the first intensive decontamination, no further bacteria and moulds could be detected.

Discussion: An important factor for the efficacy of PotoClean(®) was the age of the units and their construction. For a new generation of dental units PotoClean(®) was effective during the whole period of monitoring. For two old types of dental unit with massive biofilm development the successful decontamination needed more than 7 month.

Conclusion: The PotoClean(®) technology has resulted in even old-type turbines with intensive biofilm formation to complete decontamination. In a recent turbine design already after the first intensive decontamination with PotoClean(®) and its continuous use (1 mg Cl/L) no more contamination by bacteria and moulds were detectable.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估PotoClean(®)在日常工作中用于牙科单位供水管道再处理的去污技术。
背景:供水管道中的微生物污染和生物膜形成是牙科单位经常遇到的问题。在随机鉴定出一个细菌污染的牙科单元(310 cfu/ml)后,我们实施了本研究,以评估基于阳极氧化的PotoClean(®)技术的功效。方法:用常规低浓度永久去污(1 mg Cl/L)加PotoClean(®)强化去污(3次20 mg Cl/ml, 2小时)对3个牙单元进行7个月的疗效观察。分析了微生物污染、总氯浓度和氧化还原电位。牙科A、B组15岁,C组5岁。结果:经过3次强化净化后,A、B牙单元的细菌和霉菌数量在不到7 d的时间内即可减少。此后7个月内细菌数量再次增加,霉菌数量除个别例外为300 cfu/ml,尽管系统中不断添加PotoClean(®)(1 mg Cl/L)。再用低浓度永久消毒(1mg Cl/L) 7.5个月后,两个单元都成功去污。牙科单元C代表一个更容易去污的对象,因为先进的结构(防止水停滞)和更短的使用寿命。在净化开始时,没有细菌污染,但含有霉菌(300 cfu/ml)。在第一次密集的净化之后,已经没有进一步的细菌和霉菌被检测到。讨论:PotoClean(®)的功效的一个重要因素是单位的年龄和他们的结构。对于新一代的牙科单位,PotoClean(®)在整个监测期间都是有效的。对于两种有大量生物膜发育的旧型牙科单元,成功去污需要7个月以上。结论:PotoClean(®)技术已经导致即使是老式涡轮机也能形成密集的生物膜来完成净化。在最近的涡轮机设计中,已经在使用PotoClean(®)及其连续使用(1mg Cl/L)进行第一次密集净化后,没有更多的细菌和霉菌污染被检测到。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Evaluation of the PotoClean(®) decontamination technology for reprocessing of water supply lines in dental units during routine work. Purge- and intensive-purge decontamination of dental units contaminated with biofilm. Testing for aerobic heterotrophic bacteria allows no prediction of contamination with potentially pathogenic bacteria in the output water of dental chair units. Reprocessing of dental instruments in washer-disinfectors: does a representative test soil exist in dentistry? Impact of the "Guidelines for infection prevention in dentistry" (2006) by the Commission of Hospital Hygiene and Infection Prevention at the Robert Koch-Institute (KRINKO) on hygiene management in dental practices - analysis of a survey from 2009.
wanfang
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1